当前位置: X-MOL 学术Micron › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
A dynamic parallel image acquisition method for slide scanning process
Micron ( IF 2.4 ) Pub Date : 2023-09-09 , DOI: 10.1016/j.micron.2023.103536
Bowen Zhong 1 , Zhan Liao 1 , Xiaopeng Hao 1 , Junjie Hu 1 , Lining Sun 1
Affiliation  

Automated microscope systems have played an important role in the screening of numerous diseases. However, it is a very time-consuming process to continuously acquire images under the high magnification objective lens. This paper proposes a dynamic parallel image acquisition method, which can greatly improve image acquisition speed. Due to the relative motion between the x-y stage and the camera, some of the captured images have motion blur To this end, we also designed a motor variable speed motion curve to ensure the quality of the collected images. The experimental results show that the traditional image scanning mode needs 47.3 ms to obtain continuous microscopic images, while the dynamic parallel image acquisition method only needs 25.4 ms, which improves the acquisition speed without affecting the clarity of the acquired images.



中文翻译:

一种用于载玻片扫描过程的动态并行图像采集方法

自动显微镜系统在许多疾病的筛查中发挥了重要作用。然而,在高倍物镜下连续获取图像是一个非常耗时的过程。本文提出一种动态并行图像采集方法,可以大大提高图像采集速度。由于xy平台与相机之间的相对运动,部分采集图像存在运动模糊。为此,我们还设计了电机变速运动曲线,以保证采集图像的质量。实验结果表明,传统图像扫描方式需要47.3 ms即可获得连续显微图像,而动态并行图像采集方法仅需25.4 ms,在不影响采集图像清晰度的情况下提高了采集速度。

更新日期:2023-09-09