当前位置: X-MOL 学术Plasma Processes Polym. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Characteristics and performance of an integrated coplanar dielectric barrier discharge device for hand hygiene
Plasma Processes and Polymers ( IF 3.5 ) Pub Date : 2024-03-22 , DOI: 10.1002/ppap.202400011
Shiyun Liu 1, 2 , Haosheng Jin 1 , Yunyang Lu 1 , Li Wan 1 , Danhua Mei 1 , Zhi Fang 1
Affiliation  

A touchable plasma device based on coplanar dielectric barrier discharge (CDBD) is designed for hand hygiene. The biological safety is evaluated from the aspects of electrical and thermal properties. The peak value of current through the human body (~3.5 mA) and the highest device surface temperature (~40°C) are lower than the threshold of perception current and thermal damage temperature, respectively. The typical bacteria that are present on our hands in daily life Escherichia coli and Staphylococcus aureus are selected to verify the sterilization performance of the device. The results of human contact safety evaluation and sterilization test have demonstrated that this proposed CDBD device is reliable and user‐friendly for applications in hand hygiene.

中文翻译:

用于手卫生的集成共面介质阻挡放电装置的特性和性能

基于共面介质阻挡放电 (CDBD) 的可触摸等离子体设备专为手部卫生而设计。生物安全性从电性能和热性能方面进行评价。通过人体的电流峰值(~3.5 mA)和最高设备表面温度(~40°C)分别低于感知电流和热损伤温度的阈值。日常生活中我们手上存在的典型细菌大肠杆菌金黄色葡萄球菌选择来验证设备的灭菌性能。人体接触安全评估和灭菌测试的结果表明,所提出的 CDBD 设备对于手部卫生应用来说是可靠且用户友好的。
更新日期:2024-03-22