Effects of post-processing curing parameters and gamma irradiation on the mechanical properties of medical graded vat photopolymerization parts

Natthawut Daoset (Department of Mechanical Engineering, Faculty of Engineering at Sriracha, Kasetsart University, Chonburi, Thailand)
Samroeng Inglam (Faculty of Dentistry, Thammasat University, Pathumthani, Thailand)
Sujin Wanchat (Digital Industrial Design and Manufacturing Research Unit and Department of Mechanical Engineering, Faculty of Engineering at Sriracha, Kasetsart University, Chonburi, Thailand)
Nattapon Chantarapanich (Digital Industrial Design and Manufacturing Research Unit and Department of Mechanical Engineering, Faculty of Engineering at Sriracha, Kasetsart University, Chonburi, Thailand)

Rapid Prototyping Journal

ISSN: 1355-2546

Article publication date: 19 January 2024

Issue publication date: 23 February 2024

34

Abstract

Purpose

This paper aims to investigate the influence of post-curing temperature, post-curing time and gamma ray irradiation dose upon the tensile and compressive mechanical properties of the medical graded vat photopolymerization parts.

Design/methodology/approach

Medical graded vat photopolymerization specimens, made from photopolymer resin, were fabricated using bottom-up vat photopolymerization machine. Tensile and compressive tests were conducted to assess the mechanical properties. The specimens were categorized into uncured and post-curing groups. Temperature post-processing and/or gamma irradiation exposure were for post-curing specimens. The post-curing parameters considered included temperature levels of 50°C, 60°C and 70°C, with 1, 2, 3 and 4 h periods. For the gamma irradiation, the exposure doses were 25, 50, 75 and 100 kGy.

Findings

Post-curing improved the mechanical properties of medical graded vat photopolymerization parts for both tensile and compressive specimens. Post-curing temperature greater than 50°C or a prolonged post-curing period of more than 1 h made insignificant changes or deterioration in mechanical properties. The optimal post-curing condition was therefore a 50°C post-curing temperature with 1 h post-curing time. Exposure to gamma ray improved the compressive mechanical properties, but deteriorated tensile mechanical properties. Higher gamma irradiation doses could decrease the mechanical properties and also make the part more brittle, especially for doses more than 25 kGy.

Originality/value

The obtained results would be beneficial to the medical device manufacturer who fabricated the invasive temporary contact personalized surgical instruments by vat photopolymerization technique. In addition, it also raised awareness in excessive gamma sterilization in the medical graded vat photopolymerization parts.

Keywords

Acknowledgements

Conflict of interest: The authors declare no conflict of interest.

Citation

Daoset, N., Inglam, S., Wanchat, S. and Chantarapanich, N. (2024), "Effects of post-processing curing parameters and gamma irradiation on the mechanical properties of medical graded vat photopolymerization parts", Rapid Prototyping Journal, Vol. 30 No. 3, pp. 475-489. https://doi.org/10.1108/RPJ-07-2023-0226