An LC–MS-based designated comparison method with similar performance to the Lp(a) reference measurement procedure to guide molar Lp(a) standardization

Abstract

Background

The 2022 consensus statement of the European Atherosclerosis Society (EAS) on lipoprotein(a) (Lp(a)) recognizes the role of Lp(a) as a relevant genetically determined risk factor and recommends its measurement at least once in an individual’s lifetime. It also strongly urges that Lp(a) test results are expressed as apolipoprotein (a) (apo(a)) amount of substance in molar units and no longer in confounded Lp(a) mass units (mg/dL or mg/L). Therefore, IVD manufacturers should transition to molar units. A prerequisite for this transition is the availability of an Lp(a) Reference Measurement Procedure (RMP) that allows unequivocal molecular detection and quantification of apo(a) in Lp(a). To that end an ISO 17511:2020 compliant LC–MS based and IFCC-endorsed RMP has been established that targets proteotypic peptides of apolipoprotein(a) (apo(a)) in Lp(a). The RMP is laborious and requires highly skilled operators. To guide IVD-manufacturers of immunoassay-based Lp(a) test kits in the transition from mass to molar units, a Designated Comparison Method (DCM) has been developed and evaluated.

Methods

To assess whether the DCM provides equivalent results compared to the RMP, the procedural designs were compared and the analytical performance of DCM and RMP were first evaluated in a head-to-head comparison. Subsequently, apo(a) was quantified in 153 human clinical serum samples. Both DCM and RMP were calibrated using external native calibrators that produce results traceable to SRM2B. Measurement uncertainty (MU) was checked against predefined allowable MU.

Results

The major difference in the design of the DCM for apo(a) is the use of only one enzymatic digestion step. The analytical performance of the DCM and RMP for apo(a) is highly similar. In a direct method comparison, equivalent results were obtained with a median regression slope 0.997 of and a median bias of − 0.2 nmol/L (− 0.2%); the intermediate imprecision of the test results was within total allowable error (TEa) (CVa of 10.2% at 90 nmol/L).

Conclusions

The semi-automated, higher throughput, LC–MS-based method for Lp(a) meets the predefined analytical performance specifications and allowable MU and is hence applicable as a higher order Designated Comparison Method, which is ideally suited to guide IVD manufacturers in the transition from Lp(a) mass to molar units.

Introduction

The 2022 expert consensus by the European Atherosclerosis Society (EAS) recommends that in all individuals an assessment of lipoprotein (a) (Lp(a)) should be conducted alongside the standard serum lipid profile at least once in a lifetime [1, 2]. While Lp(a) as a risk factor for cardiovascular disease has long been controversial, these recommendations are now based on updated evidence for a causal continuous association in different ethnic groups between Lp(a) concentrations and adverse cardiovascular outcomes [1]. The EAS expert consensus moreover advises that Lp(a) test results are expressed as apo(a) amount of substance in molar units -due to extreme apo(a) heterogeneity- and no longer in confounded Lp(a) mass units (mg/dL or mg/L). With these recommendations, as well as the promise of Lp(a) lowering drugs [1, 3,4,5,6], the need for accurate quantification of Lp(a) in molar units is now urgent.

Currently, many IVD manufacturers market kits that provide results in mass units even though standardization and expression of Lp(a) measurements in molar units have been recommended for > 2 decades [1, 7,8,9]. Flawed Lp(a) results should be overcome and therefore IVD manufacturers must transition from mass units to molar units. A prerequisite for this transition is the availability of an Lp(a) Reference Measurement Procedure (RMP) that allows unequivocal molecular detection and quantification of apolipoprotein(a) (apo(a)) in Lp(a). The International Federation for Clinical Chemistry and Laboratory Medicine (IFCC) working group for standardization of apolipoproteins by mass spectrometry (IFCC WG APO-MS) set out a conceptual approach for such an RMP [8], and recently reported on an ISO 17511:2020 compliant liquid chromatography – mass spectrometry (LC–MS) based and IFCC-endorsed RMP [10]. The RMP quantifies proteotypic peptides of apo (a) which is the specific apolipoprotein and a major structural component of lipoprotein(a) [11]. Importantly, apo(a) is a highly heterogenic protein that manifests in various proteoforms, due to the KIV2 polymorphisms and post translational modifications [10, 12]. Moreover, it is present in serum in a wide concentration range. As MS allows for quantification of molecularly defined peptides, the apo(a) peptides were selected to accurately reflect the intended measurand (apo(a) amount of substance in molar units).

A RMP is a higher order measurement procedure in the metrological traceability chain, as outlined in ISO 17511:2020 [13]. Therefore, the method has to be of high quality and execution of the procedure should take place according to stringent quality measures as outlined in ISO 15195:2018 [14]. Such high quality and confidence in measurement accuracy, however, comes at the cost of ease of use of the procedure. To facilitate the transition from mass to molar units, a Designated Comparison Method (DCM) would be beneficial. To that end, a semi-automated LC–MS based method for quantitation of apo(a) in molar units was developed; the method is easier to operate and comprises only a single proteolytic digestion step.

To ensure the method is suited for its intended purpose, the DCM should produce equivalent results and have a comparable measurement uncertainty as the RMP. Its total error budget should only be a fraction of the allowable error budget (TEa), as deduced from biological variation. This allowable error is set at the level of the end user, but is divided over the various stakeholders/methods in the traceability chain. We therefore propose that the DCM may not consume more than 1/2 of the total allowable error (TEa) budget. For trueness, the correlation coefficient of a comparison between the two methods should be > 0.975, and the slope of the comparison may not significantly deviate from 1.000 under the used similar external calibration procedure. In this study, we performed a head-to-head comparison between the RMP and the DCM, both at the levels of their design and analytical performance, to ensure results generated with the DCM can be used to guide IVD manufacturers in their transition towards nmol/L units.

Materials and methods

This section briefly describes the semi-automated DCM compared to the previously published RMP [15]. Subsequently, the method comparison is described. Common solvents and consumables are described at the start of this section.

Materials

Sequencing grade modified trypsin (V5111) and mass spectrometry grade Lys-C (VA1170) were acquired from Promega (Leiden, The Netherlands). LC–MS grade methanol (MeOH) was purchased from Biosolve (Valkenswaard, The Netherlands) and formic acid (FA, ≥ 99% purity) was purchased from Avantor/VWR (Radnor, PA, USA). Ammonium bicarbonate (ABC) was purchased from Fluka (Landsmeer, the Netherlands) while Iodoacetamide (IAM), Sodium Deoxycholate (DOC) and tris(hydroxymethyl)aminomethane (TRIS), pH 8.1, were purchased from Sigma-Aldrich (Zwijndrecht, The Netherlands). Tris(2-carboxyethyl)phosphine (TCEP) was purchased from Thermofisher Scientific (Waltham, MA). Synthetic and stable isotope labelled peptides were produced by our in-house facility. The assessment of their purity was conducted using LC with UV detection and matrix assisted laser desorption/ionization (MALDI)-MS. Deidentified serum samples for calibration and quality control were procured from MCA Laboratory (Winterswijk, The Netherlands).

VACUETTE® Secondary 13 × 75 mm MULTIPLEX PET tubes (459011), U-bottom (650201), V-bottom 96 well microplates (651201), plate lids (656102), 15 mL Cellstar tubes (188271) and 50 mL Cellstar tubes (227261) were from Greiner bio-one (Alphen aan den Rijn, the Netherlands). Skirted (0030.128.680) and semi-skirted (0030.128.613) 96-well PCR plates were from Eppendorf (Nijmegen, the Netherlands). Bioke (Leiden, the Netherlands) supplied gas permeable moisture barrier seals (4ti-0516/96) and pierceable seals (4ti-0566). 250 µL Tips for Bravo (19477–002) were from Agilent Technologies (Santa Clara, CA, USA). and Oasis Prime HLB µElution Plates (186008052) were from Waters Corporation (Milford, PA, USA).

Semi-automated LC–MS method for the quantitation of apo(a)

Apo(a) quantification was implemented in our already existing test for multiplexed apolipoprotein quantification and was conducted according to the standard operating procedure for sample preparation and tryptic digestion, as outlined previously [15, 16]. In short, Samples were 20 × diluted in 100 mmol/L TRIS, followed by protein solubilization and cysteine reduction using DOC and TCEP at 56 °C. Cysteine thiols were methylated by IAM at room temperature in the dark, immediately followed by trypsin digestion at a protein:enzyme ratio of 35:1 (w/w) at 37 °C. After 3 h, the reaction is quenched and peptides formed are enriched using Oasis HLB SPE, eluting the peptides with 55% MeOH. The workflow is semi-automated on a 96-channel BRAVO automated liquid handling platform (Agilent Technologies). The LC–MS system consisted of a 1290 multisampler (G7167B), and 1290 high speed analytical pump (G7120A) and 1290 multicolumn thermostat (G7116B), coupled to either a 6495A or 6495C triple quadrupole mass spectrometer. The instrument was operated in positive mode, in dynamic MRM mode with a cycle time of 500 ms. For apo(a), two peptides were monitored, each with three transitions. A method-specific system suitability test, comprising measurement of a mixture of synthetic and SIL peptides was performed along with each analytical run. An overview of the DCM and RMP method similarities and disparities can be found in Table 1.

Table 1 Head-to-Head comparison of the fundamental characteristics defining of the IFCC-endorsed Reference Measurement Procedure (RMP) and the Designated Comparison Method (DCM) for apo(a)

Reference Measurement Procedure for the quantitation of apo(a)

Apolipoprotein quantification was conducted according to the standard operating procedure, as outlined previously [10]. In short, samples were 20 × diluted in 100 mmol/L ABC. Proteins were solubilized by DOC and cysteines were reduced using TCEP at 56 °C. Cysteine thiols were methylated by IAM at room temperature in the dark, immediately followed by pre-digestion using Lys-C for 1 h at 37 °C at a protein:enzyme ratio of 700:1 (w/w). Then trypsin was added at a protein:enzyme ratio of 47:1 (w/w) and samples were further digested at 37 °C for 3 h. The reaction was quenched, and Oasis HLB SPE was employed for peptide enrichment, eluting the peptides with 80% MeOH. The workflow was performed manually. The LC–MS system consisted of a 1290 multisampler (G7167B), and 1290 high speed analytical pump (G7120A), and 1290 multicolumn thermostat (G7116B), coupled to a 6495A triple quadrupole mass spectrometer. The instrument was operated in positive mode, in dynamic MRM mode with a cycle time of 500 ms. For apo(a), three peptides were monitored, each with three transitions. A method-specific system suitability test, comprising measurement of a mixture of synthetic and SIL peptides was performed along with each analytical run. An overview of the method can be found in Table 1.

Comparison between RMP and semi-automated Designated Comparison Method for apo(a)

A head-to-head comparison between the RMP and the DCM was performed based on their standard operating procedures and equivalence and analytical performance were evaluated according to CLSI EP-15. Moreover, analytical imprecision, as assessed through both a CLSI EP-15 protocol (measurement of five human serum samples in quintuplicate on five different days (total n = 25 per sample) and long-term IQC monitoring, were compared.

Deidentified Serum samples that were left over from general clinical chemistry analysis were collected from 153 human donors. In short, phlebotomy was conducted in Becton Dickinson (BD) serum gel tubes (367957) and blood was allowed to clot at room temperature for 30 min, followed by centrifugation at 3000 g for 8 min. The donors provided broad consent for the use of their deidentified biomaterial. An aliquot containing 250 μl was prepared immediately after centrifugation and stored at – 20 ℃ for up to 3 weeks. First, the DCM was performed in eight batches. After refreezing at − 20 °C for 3 to 7 months, the samples were analyzed with the RMP in two batches.

For data analysis, apo(a) concentrations below the limit of quantitation of < 3.8 nmol/L were excluded from method comparison analysis. Concentrations obtained by the DCM were plotted against the RMP results and linear regression was performed.

Results

Head-to-head comparison of the methods

Results of the head-to-head comparison of the RMP and the DCM are summarized in Table 1. While the procedures are highly similar and both based on the principle of bottom-up proteomics, differences between the procedure the facilitate ease of use of the DCM have been made. Specific differences between the methods are alternative peptide and transition selection, the use of a single calibrator (RMP) vs the use of five calibrators (DCM), the pre-digestion (RMP) and semi-automation (DCM). Each of these aspects are further elaborated below.

For the RMP, three proteotypic peptides are selected for quantitation: GISSTVTGR (kringle 5), LFLEPTQADIALLK (protease domain) and TPENYPNAGLTR (kringle 9). For the DCM, only two peptides are monitored: GISSTVTGR and LFLEPTQADIALLK. In both procedures three transitions are monitored per peptide; for peptide GISSTVTGR, the three transitions overlap, while for peptide LFLEPTQADIALLK, the b2 fragment is monitored in the RMP and the b3 fragment is monitored in the DCM (Table 2). The calibration of both the RMP and the DCM is in nmol/L and is currently based on native human serum calibrators that are indirectly traceable to WHO-IFCC reference material SRM2B. For the RMP a single point calibration is used (93.4 nmol/L), while 5 calibrators (17.1–270.7 nmol/L) are used in the DCM.

Table 2 Transitions monitored for the quantitation of apo(a) in both the RMP and the DCM

Minor differences are present in the reagents used in the sample preparation: A 100 mM ABC buffer (pH 8.1) is used for the RMP, while a 100 mM TRIS buffer (pH 8.1) is used in the DCM. 1.15 mmol/L TCEP was used in the RMP as well as 0.40% (v/v) DOC and 4.6 mmol/L IAM. For the DCM 0.5 mM TCEP, 0.523% (v/v) DOC and 4.6 mmol/L IAM was used. A pre-digestion step using LysC is performed prior to trypsin digestion (1:47, w/w) in the RMP, while in the DCM trypsin digestion (1:35, w/w) is started immediately. SPE of the two procedures is performed on the HLB stationary phase, with elution using 80% MeOH in the RMP compared to 55% MeOH in the DCM. A major difference in the execution of the sample preparation is the semi-automation of the DCM, which is not yet implemented for the RMP.

Analytical performance: intermediate imprecision

While it is anticipated that the procedural differences between the RMS and the DCM will not cause deviating results, the analytical performances of both tests are compared. Medical tests should be fit-for-purpose. Therefore, their analytical performance specifications should be predefined. The allowable error budgets are set according to the Milan hierarchy [18], based ideally on clinical outcome, alternatively on biological variation data, and in other cases based on state-of-the-art. To the best of our knowledge, no data is available to defer error budgets for apo(a) based on clinical outcome studies; therefore, analytical performance specifications and total allowable error (TEa) based on biological variation are used. However, calculations typically rely on constant biological variation, independent of the analytes concentration. Yet, Lp(a) is a heterogenous particle containing the highly variable protein apo(a). This variability extends to both the concentration and structure of apo(a), including kringle IV-2 size polymorphism, as well as N- and O- glycosylation, leading to 1000-fold differences between individuals [12, 19,20,21]. The Lp(a) distribution is also racially dependent and does not follow a Gaussian curve for most of the races studied [21, 22]. Consequently, the intra- and inter-individual variation vary over the range of the Lp(a) concentrations. Indeed, a logarithmic intra-individual biological variation (CVi) was observed [23]. In error budget calculations, the logarithmic CVi can be considered. Measurements in Cobbaert et al. [23] were performed using a former state-of-the-art Lp(a) test, not yet standardized to the former Reference Measurement System (RMS) [23, 24]. However, assuming a (linear) conversion factor of 300 mg/L = 90 nmol/L, and a constant Inter-individual Coefficient of Variation (CVg) of 18.1%, a concentration dependent minimum TEa can be deduced (Fig. 1, grey line). The total error budget must be divided over the different stakeholders of the traceability chain and as such, the proportional error budget for the RMP was set at 50% (Fig. 1, red line).

Fig. 1
figure 1

Comparison of imprecision of RMP and DCM based on CLSI EP-15 protocol results. Tests should fulfill predefined analytical performance specifications. For apo(a), the minimum total allowable error (TEa) is concentration dependent (grey line). To divide the error budget over the stakeholders of the metrological traceability chain, ½ of the TEa is allocated for the RMP, and thus ideally also the DCM (red line). EP-15 results are plotted as blue dots (RMP) or green dots (DCM)

To compare the analytical imprecision of the DCM with the RMP, the results of CLSI EP-15 protocols for evaluation of analytical imprecision (Table 3) are plotted in Fig. 1 (dots). For quantifying peptide LFLEPTQADIALLK, the analytical imprecision of the DCM fulfills the analytical performance specifications, while for qualifying peptide GISSTVTGR the imprecision is close to fulfillment of the analytical performance specifications. However, as the peptide is included for confirmation purposes, the criteria can be loosened.

Table 3 Results of an EP-15 analytical imprecision verification of both the RMP and the DCM

The long-term imprecision has also previously been monitored for both the RMP and the DCM. For the RMP, a native human serum IQC sample containing 44 nmol/L apo(a) yielded CVs of 9.0% and 10.0% for peptides GISSTVTGR and LFLEPTQADIALLK, respectively over a 2-year period, 84 measurements and 34 96-well batches [10]. Similar results have been obtained for the DCM, where a native human serum sample containing 102 nmol/L apo(a) yielded a CV of 7.3% for peptide LFLEPTQADIALLK over a 2-year period [16, 17]. Based on the comparison of both the CLSI EP-15 results as well as the long-term robustness, it can be concluded that the analytical performance of the semi-automated DCM is comparable to that of the RMP (Table 4)

Table 4 Summary statistics of IQC performance during method comparison

Data validity for the method comparison

To define whether results obtained with the DCM are exchangeable with results obtained with the RMP, a method comparison was performed on 153 native human serum samples, each measured individually. First, measures of data validity (IQC, and assessment of intrinsic metadata) are presented, followed by the method comparison. The RMP results were generated in two batches, containing bilevel IQC (total n = 5). Average concentrations were 41 nmol/L and 7 nmol/L for QC1 and QC2 for peptide LFLEPTQADIALLK, with CVs of 5.2% and 6.9%. For the DCM, apo(a) was quantified in seven batches, each containing bilevel IQC (total n = 21). Average concentrations were 40 nmol/L and 8 nmol/L for QC1 and QC2 for peptide LFLEPTQADIALLK, with CVs of 5.2% and 3.3%.

To further assess quality of the individual sample results, ion ratios between quantifying and qualifying transitions, internal standard areas and interpeptide agreements were monitored for both the RMP as well as the DCM. Ion ratios are affected by MS instrument settings, and the DCM was performed on four different LC–MS systems. Therefore, summary statistics of the ion ratios were calculated for each measurement batch individually. Overall, the ion ratios are consistent between endogenous peptides and internal standard peptides and CVs range between 7.3% and 12.8% for peptide LFLEPTQADIALLK of the RMP and 5.3% and 26% for the DCM (Table 5), indicating consistent measurements throughout each batch. The ion ratios of the individual samples are plotted in Fig. 2; no individual outliers were observed, suggesting that results are not confounded by interferences.

Table 5 Summary statistics of ion ratio monitoring
Fig. 2
figure 2

Ion ratio monitoring of the results obtained with the RMP (top panel) and the DCM (lower panel). Ion ratios were defined as the ratio between the quantifying transition and each of the qualifying transitions (Table 2). Dotted lines indicate different measurement batches, which may have occurred at different instruments (n = 4) for the DCM, resulting in inter-batch variability. However, within a batch, consistent ion ratios are observed

A third measure of measurement quality is the stability of the internal standard. Summary statistics (mean and CV) of the signal intensities of the internal standard peptides can be found in Table 6 and Fig. 3. In Fig. 3, the observed differences in batches can be attributed to the use of multiple mass spectrometers, specifically the 6495A and 6495C triple quadrupole instruments. However, for the RMP, the variation was low, with CVs ranging between 6.9% and 10.3%. Similar CVs were observed for the DCM.

Table 6 Summary statistics of internal standard area monitoring
Fig. 3
figure 3

Internal standard area monitoring of the results obtained with the RMP (top panel) and DCM (lower panel). Areas from each of the measurements is plotted, and the various different batches are indicated by vertical dotted lines

A final tool to assess data validity is interpeptide comparison (Fig. 4). As each of the peptides monitored should represent the same protein, results of the peptides should provide equivalent results, within analytical variation. Comparisons were made for both the RMP and the DCM and Passing-Bablok regression analysis revealed slopes of 0.989 and 0.950 between peptide LFLEPTQADIALLK and peptides GISSTVTGR and TPENYPNAGLTR, respectively for the RMP and 1.04 between peptide LFLEPTQADIALLK and GISSTVTGR for the DCM (Table 7). Correlation coefficients were also as expected, 0.987 and 0.984 for the RMP and 0.985 for the DCM. Overall, assessment of these quality measures indicated that the performance of the DCM generated valid and robust metadata, comparable to those of the RMP.

Fig. 4
figure 4

Interpeptide comparisons for both the RMP and the DCM. Passing-Bablok regression is indicated in the regression plot (red line), while the black line indicates the line of identity. Red dotted line in the bias plots indicates median (percent) bias

Table 7 Results of interpeptide comparison by Passing-Bablok non-parametric regression analysis

Method comparison

A direct method comparison was performed between the RMP and the DCM. The LoQ of the RMP is 3.8 nmol/L and results below this threshold were removed from the analysis per peptide. Ordinary least squares linear regression was employed for the method comparison, as the RMP, which is the anchor point, was included on the x-axis. Results of the method comparisons for each of the peptides are summarized in Table 8 and shown in Fig. 5. Median slope for the comparisons was 0.997, with slopes ranging between 0.944 and 1.052. Negligible and non-significant intercepts were observed. Biases ranged between -2.2 nmol/L and 3.5 nmol/L, with a median bias of 0.2 nmol/L (0.2%). Correlation coefficients are all ≥ 0.975 except for the comparison between RMP peptide TPENYPNAGLTR and DCM peptide GISSTVTGR (r = 0.969). However, peptide TPENYPNAGLTR is included as qualifying peptide, only to confirm quantitative results. The observed variation between the RMP and the DCM are within expected analytical variation, therefore the performance of the RMP and the DCM is equivalent.

Table 8 Results of method comparison
Fig. 5
figure 5

Method comparisons between results obtained with the Reference Measurement Procedure (RMP) and the Designated Comparison Method (DCM). Each of the RMP peptides is compared to each of the DCM peptides. Linear regression is used, as the X-axis represents results from the RMP, and is indicated in the regression plots (red line), while the black lines indicate the line of identity. Red dotted line in the bias plots indicated median (percent) bias

Discussion

With the inclusion of Lp(a) as a relevant genetic risk factor in clinical guidelines on cardiovascular risk management, and Lp(a) lowering therapies in phase III clinical studies, accurate measurement of Lp(a) in medical laboratories is more urgent than ever. As Lp(a) and its characteristic apolipoprotein apo(a) are highly heterogeneous, measurements of apo(a) in molar units are recommended. To ensure an unequivocally defined measurand, a new RMS based on SI-traceable primary calibrators and an LC–MS based RMP is in preparation by the IFCC WG APO-MS [8, 25]. However, until the full RMS including the calibration is in place, a method is needed to pave the way for the transition from mass units to molar units with IVD manufacturers.

To this end, we developed a semi-automated MS-based DCM for apo(a), with equivalent performance compared to the more laborious RMP. The DCM is based on the quantitative bottom-up proteomics strategy in which peptides are quantified as surrogates of proteins. Therefore, stringent peptide selection is imperative [17], as has been described for the RMP [10]. In the RMP three proteotypic peptides are quantified: GISSTVTGR, LFLEPTQADIALLK and TPENYPNAGLTR. Peptide TPENYPNAGLTR may be prone to a genetic variant, which occurs in ~ 6% of individuals of African descent and may only serve as qualifying peptide to confirm obtained quantitative results [10]. For the DCM, monitoring of two peptides suffices and the best performing peptides in terms of precision (GISSTVTGR and LFLEPTQADIALLK) have been included in the DCM.

Analytical specificity of the method is (among others) achieved through the selection of specific transitions. For the RMP, common transitions were selected that could be monitored in all three calibration laboratories of the Lp(a) calibration network, using different LC/MS-instruments. However, the DCM was developed in a single calibration laboratory. Therefore, for the RMP the analytically less specific b2 fragments were included as qualifying fragments, whereas the b3 fragments could be selected for the DCM (Table 2).

The native human serum-based calibration used in the RMP has a temporary character, as an SI-traceable calibration is envisioned. Therefore, the RMP makes use of only a single calibrator for apo(a). Contrarily, the DCM employs five native human serum calibrators. The calibrators of both the RMP and the DCM were indirectly value assigned for their apo(a) concentration using an immunoassay traceable to WHO-IFCC reference material SRM2B.

A third major difference between the RMP and the DCM is the digestion procedure. For the RMP, peptide-based calibration is envisioned to achieve traceability to the International System of measures [8], which requires complete, equimolar digestion. Moreover, the RMP is a multiplexed procedure, not only intended for quantitation of apo(a), but also other apolipoproteins. To enhance the digestion efficiency and reach stable digestion plateaus for all the peptides monitored in the RMP, a pre-digestion using Lys-C was introduced [10]. This step is, however, costly, both timewise and monetarily. As the DCM will not use primary reference materials for calibration, but native human serum calibrators (secondary reference materials), only stable digestion is required, as long as the calibrators behave the same compared to the human serum samples. Intermediate imprecision of the RMP and the DCM are equivalent and fulfill the predefined analytical performance specifications of 50% of the TEa (ranging between 21.3% and 7.8% at Lp(a) concentrations of 5.0 nmol/L and 344 nmol/L, respectively). Results of the method comparison also indicate equivalence of results. Therefore, it may be concluded that the method may be used as a higher order DCM for the apo(a) RMP.

An essential step for the increased throughput in the DCM is the application of a liquid handling robot for semi-automation of the sample preparation method. As we transition from the old RMS to the new RMS, the IVD-industry will be directed through the utilization of the DCM for apo(a). LC–MS technology is, currently, not yet available on consolidated and robust clinical chemistry analyzers. Therefore, sample preparation must be performed in ‘batch-mode’. Batches of 96 are ideally suited for semi-automation using liquid handlers, providing more consistent test results combined with a more facile and less error-prone procedure. While the RMP was not yet implemented on a liquid handling platform, the semi-automated DCM is, which enables higher sample throughput.

Both the RMP and the DCM are LC–MS based procedures. LC–MS provides significant advantages of traditional immunoassays for the quantitation of heterogeneous proteins [26]. Specifically, the detection method allows quantitation of molecularly defined measurands [17], is antibody independent, and may generate intrinsic metadata to determine validity of individual results [27]. The measurands as defined in the DCM are molar concentrations of peptides GISSTVTGR and LFLEPTQADIALLK, reflecting apo(a) in human serum, and are by definition independent of apo(a) KIV-2 repeats. An important disadvantage of immunoassays is their vulnerability for interferences generated by non-specific binding or masking of the target protein. In the RMP and DCM LC–MS procedures both IQC and intrinsic metadata were assessed to ensure individual results generated are correct. In particular, ion ratios between quantifying and qualifying transitions, stability of the internal standard signal and interpeptide agreements may reveal variation, and monitoring of these parameters is therefore recommended (CLSI C64)[28], and was performed.

In the direct method comparison performed between the RMP and the DCM equivalent results were obtained. Therefore, the DCM may be used as a more facile method compared to the RMP in the establishment of standardization of Lp(a) tests.

Current Lp(a) tests may or may not be traceable to the previous RMS, that consisted of a reference material that is not SI-traceable, but value assigned in nmol/L (SRM2B), and a KIV-2 independent ELISA-based RMP (Fig. 6) [29, 30]. However, this RMS is no longer available, as the material has run out of stock and the RMP is no longer operational. The new LC–MS based RMS is still in development, and therefore a 2-phased process for Lp(a) re-standardization is advocated by the IFCC WG APO-MS: first, the transition from mass to molar units should be accomplished, to overcome the unavailability of an operational RMS for apo(a). Secondly, the final step for Lp(a) re-standardization should be realized when the entire MS-based RMS is in place, including the peptide-based calibration. As the results of the RMP and the DCM are equivalent, the DCM can be used as a provisional anchor to guide IVD-manufacturers with Lp(a) re-standardization to molar units. Specifically, a method comparison program is being built in which well-characterized samples are quantified using both the DCM and the manufacturer’s method. The samples will cover a wide Lp(a) concentration range, as well as a variety of known apo(a) KIV-2 repeats. Through evaluation of the obtained data, manufacturers of Lp(a) immunoassays will be able to obtain data on three important aspects (Fig. 6): 1. Recommendations on their current level of traceability to previous reference material SRM2B in nmol/L, including the level of KIV-2 dependency of their test. 2. Recommendations on the effects of sample dilution in samples containing high apo(a) concentrations that are above the linear measuring range. And 3. Information on the expected implications of the transition from the previous ELISA-based RMS to the new LC–MS based RMS.

Fig. 6
figure 6

Schematic representation of the transition from the former ELISA-based Reference Measurement System (NWLRC, Washington, USA) to an MS-based Reference Measurement System [8] that is in development. The here presented Designated Comparison Method (DCM) will have a key role in the transition phase, in which manufacturers can maintain SRM2B traceability, while abolishing inaccuracy due to multiple misinterpretations around the Lp(a) metrics

Conclusion

We here present a semi-automated LC–MS based DCM for apo(a)/Lp(a) that is apo(a) KIV-2 independent, fulfills predefined analytical performance specifications and provides equivalent results compared to the IFCC-endorsed apo(a) RMP [10]. The Lp(a) DCM will be utilized to guide IVD-manufacturers to make the necessary transition from flawed mass units to molar units. While the transition to well defined apo(a) measurands and their expression in molar units is a long and challenging road, the apo(a) transition from mass to molar is an essential first step to document the relation between former and new RMP, to solve the metrological misunderstandings around Lp(a) metrics and to enable more refined cardiovascular patient management.

Availability of data and materials

The datasets generated and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ABC:

Ammonium bicarbonate

Apo:

Apolipoprotein

APS:

Analytical Performance Specifications

BD:

Becton Dickinson

CLSI:

Clinical and Laboratory Standards Institute

CV:

Coefficient of Variation

CVi:

Intra-individual Coefficient of Variation

CVg:

Inter-individual Coefficient of Variation

DCM :

Designated Comparison Method

DOC:

Sodium Deoxycholate

ELISA:

Enzyme-Linked Immunosorbent Assay

FA :

Formic Acid

HLB :

Hydrophilic-Lipophilic Balance

IAM:

Iodoacetamide

IQC:

Internal Quality Control

IFCC:

International Federation of Clinical Chemistry and Laboratory Medicine

IFCC WG APO-MS:

International Federation of Clinical Chemistry and Laboratory Medicine Working Group for Standardization of Apolipoproteins by Mass Spectrometry

ISO:

International Organization for Standardization

IVD:

In vitro diagnostics

KIV-2:

Kringle IV type 2

LC-MS:

Liquid Chromatography—Mass Spectrometry

Lp(a) :

Lipoprotein(a)

MeOH :

Methanol

RMP:

Reference Measurement Procedure

RMS:

Reference Measurement System

SI:

International System of Units

SIL :

Stable Isotope Labeling

SRM2B:

Standard Reference Material 2B

TCEP:

Tris(2-carboxyethyl)phosphine

TEa:

Total Allowable Error

TRIS:

Tris(hydroxymethyl)aminomethane

References

  1. Kronenberg F, Mora S, Stroes ESG, et al. Lipoprotein(a) in atherosclerotic cardiovascular disease and aortic stenosis: a European Atherosclerosis Society consensus statement. Eur Heart J. 2022;43(39):3925–46.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Kronenberg F, Mora S, Stroes ESG, et al. Frequent questions and responses on the 2022 lipoprotein(a) consensus statement of the European Atherosclerosis Society. Atherosclerosis. 2023;374:107–20.

    Article  CAS  PubMed  Google Scholar 

  3. Tsimikas S, Karwatowska-Prokopczuk E, Gouni-Berthold I, et al. Lipoprotein(a) reduction in persons with cardiovascular disease. N Engl J Med. 2020;382(3):244–55.

    Article  CAS  PubMed  Google Scholar 

  4. Koren MJ, Moriarty PM, Baum SJ, et al. Preclinical development and phase 1 trial of a novel siRNA targeting lipoprotein(a). Nat Med. 2022;28(1):96–103.

    Article  CAS  PubMed  Google Scholar 

  5. Nissen SE, Wolski K, Balog C, et al. Single ascending dose study of a short interfering RNA targeting lipoprotein(a) production in individuals with elevated plasma lipoprotein(a) levels. JAMA. 2022;327(17):1679–87.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Nicholls SJ, Nissen SE, Fleming C, et al. Muvalaplin, an oral small molecule inhibitor of lipoprotein(a) formation: a randomized clinical trial. JAMA. 2023. https://doi.org/10.1001/jama.2023.16503.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Tsimikas S, Fazio S, Viney NJ, et al. Relationship of lipoprotein(a) molar concentrations and mass according to lipoprotein(a) thresholds and apolipoprotein(a) isoform size. J Clin Lipidol. 2018;12(5):1313–23.

    Article  PubMed  Google Scholar 

  8. Cobbaert CM, Althaus H, Begcevic Brkovic I, et al. Towards an SI-traceable reference measurement system for seven serum apolipoproteins using bottom-Up quantitative proteomics: conceptual approach enabled by cross-disciplinary/cross-sector collaboration. Clin Chem. 2021;67(3):478–89.

    Article  PubMed  Google Scholar 

  9. Tsimikas S, Fazio S, Ferdinand KC, et al. NHLBI working group recommendations to reduce lipoprotein(a)-mediated risk of cardiovascular disease and aortic stenosis. J Am Coll Cardiol. 2018;71(2):177–92.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Ruhaak LR, Romijn F, Begcevic Brkovic I, et al. Development of an LC-MRM-MS-based candidate reference measurement procedure for standardization of serum apolipoprotein (a) tests. Clin Chem. 2023;69(3):251–61.

    Article  PubMed  Google Scholar 

  11. Berg K. A new serum type system in man–the Lp system. Acta Pathol Microbiol Scand. 1963;59:369–82.

    Article  CAS  PubMed  Google Scholar 

  12. Ruhaak LR, Cobbaert CM. Quantifying apolipoprotein(a) in the era of proteoforms and precision medicine. Clin Chim Acta. 2020;511:260–8.

    Article  CAS  PubMed  Google Scholar 

  13. ISO (International Organization for Standardization) 17511;2020. Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples. https://www.iso.org/standard/69984.html.

  14. ISO. (International, Organization, for, Standardization) 15195;2018. Laboratory medicine — Requirements for the competence of calibration laboratories using reference measurement procedures. https://www.iso.org/standard/69824.html.

  15. van den Broek I, Romijn FP, Nouta J, et al. Automated multiplex LC-MS/MS assay for quantifying serum apolipoproteins A-I, B, C-I, C-II, C-III, and E with qualitative apolipoprotein E phenotyping. Clin Chem. 2016;62(1):188–97.

    Article  PubMed  Google Scholar 

  16. Ruhaak LR, Smit NPM, Romijn F, et al. Robust and accurate 2-year performance of a quantitative mass spectrometry-based apolipoprotein test in a clinical chemistry laboratory. Clin Chem. 2018;64(4):747–9.

    Article  CAS  PubMed  Google Scholar 

  17. Smit NPM, Ruhaak LR, Romijn F, et al. The time has come for quantitative protein mass spectrometry tests that target unmet clinical needs. J Am Soc Mass Spectrom. 2021;32(3):636–47.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Sandberg S, Fraser CG, Horvath AR, et al. Defining analytical performance specifications: consensus statement from the 1st strategic conference of the european federation of clinical chemistry and laboratory medicine. Clin Chem Lab Med. 2015;53(6):833–5.

    Article  CAS  PubMed  Google Scholar 

  19. Subramanian SP, Gundry RL. The known unknowns of apolipoprotein glycosylation in health and disease. iScience. 2022;25(9):105031.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  20. Reyes-Soffer G, Ginsberg HN, Berglund L, et al. Lipoprotein(a): a genetically determined, causal, and prevalent risk factor for atherosclerotic cardiovascular disease: a scientific statement from the american heart association. Arterioscler Thromb Vasc Biol. 2022;42(1):e48–60.

    Article  CAS  PubMed  Google Scholar 

  21. Kronenberg F. Human genetics and the causal role of lipoprotein(a) for various diseases. Cardiovasc Drugs Ther. 2016;30(1):87–100.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Guan W, Cao J, Steffen BT, et al. Race is a key variable in assigning lipoprotein(a) cutoff values for coronary heart disease risk assessment: the Multi-Ethnic Study of Atherosclerosis. Arterioscler Thromb Vasc Biol. 2015;35(4):996–1001.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Cobbaert C, Arentsen JC, Mulder P, et al. Significance of various parameters derived from biological variability of lipoprotein(a), homocysteine, cysteine, and total antioxidant status. Clin Chem. 1997;43(10):1958–64.

    Article  CAS  PubMed  Google Scholar 

  24. Marcovina SM, Albers JJ. Lipoprotein (a) measurements for clinical application. J Lipid Res. 2016;57(4):526–37.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  25. Apolipoproteins by Mass Spectrometry (WG-APO MS) [cited 2023 11–09–2023]. https://ifcc.org/ifcc-scientific-division/sd-working-groups/wg-apo-ms/

  26. Hoofnagle AN, Wener MH. The fundamental flaws of immunoassays and potential solutions using tandem mass spectrometry. J Immunol Methods. 2009;347(1–2):3–11.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Vogeser M, Schuster C, Rockwood AL. A proposal to standardize the description of LC-MS-based measurement methods in laboratory medicine. Clin Mass Spectrom. 2019;13:36–8.

    Article  PubMed  PubMed Central  Google Scholar 

  28. CLSI. Quantitative measurement of proteins and peptides by mass spectrometry. CLSI guideline C64. 1st ed. Clinical and Laboratory Standards Institute; 2021.

    Google Scholar 

  29. Dati F, Tate JR, Marcovina SM, et al. First WHO/IFCC international reference reagent for lipoprotein(a) for immunoassay–Lp(a) SRM 2B. Clin Chem Lab Med. 2004;42(6):670–6.

    Article  CAS  PubMed  Google Scholar 

  30. Tate JR, Berg K, Couderc R, et al. International federation of clinical chemistry and laboratory medicine (IFCC) standardization project for the measurement of lipoprotein(a). Phase 2: selection and properties of a proposed secondary reference material for lipoprotein(a). Clin Chem Lab Med. 1999;37(10):949–58.

    Article  CAS  PubMed  Google Scholar 

Download references

Funding

No external funding was received.

Author information

Authors and Affiliations

Authors

Contributions

NMD and LRR performed data analysis and drafted the manuscript. CMC and LRR conceptualized and designed the study and revised the manuscript. MMP, NPS and FPR performed data collection.

Corresponding author

Correspondence to Christa M. Cobbaert.

Ethics declarations

Ethics approval and consent for participate

The use of deidentified patient materials for method comparison purposes is not subject to METC approval. Broad consent was obtained from all participants indicating they have no objections to use of their deidentified materials.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions