A 49-year-old inoperable male presented for treatment due to severe tricuspid regurgitation. The heart team opted to implant a TricValve (Unimedtec) considering the challenging: impacted pacemaker lead, esophageal stenosis, and a small superior vena cava (Fig. 1, Supplementary Fig. 1).
During the procedure, a 27-Fr sheath was inserted into the right femoral vein, and a 21-mm SVC valve delivery system was introduced (Panels A). However, as the system advanced, its nose cone unintentionally pushed two pacemaker leads into the right brachiocephalic vein (RBV), causing a blockage at the confluence of the brachiocephalic veins. This resulted in the pacemaker leads being carried alongside the delivery system, creating a compounded obstruction (Panel B). To solve this, a goose snare catheter was positioned to grasp and pull the pacemaker leads inferiorly (Panel C), preventing their re-entry into the RBV. However, while the SVC delivery system was trying to reintroduce to RBV, causing increased upward tension on the ventricular electrode (Panel D), the snare system simultaneously exerted more force to pull both electrodes downward, resulting in the dislocation of the atrial electrode (Panel E). After successfully deploying the SVC valve (Panel F), the team repositioned the dislocated atrial electrode (Panels G and H) and implanted a 31-mm IVC valve. Angiography confirmed correct valve placement and function (Panel I). A new atrial electrode was then placed from left brachiocephalic vein, through the cell of the SVC valve and secured in the right atrial wall (Supplementary Fig. 2).
Post-procedure, the patient had functioning SVC and IVC valves, no pacemaker issues, and was discharged in four days (Video 1).
Data availability
Data can be required from corresponding author if needed.
Funding
Open access funding provided by University of Zurich
Author information
Authors and Affiliations
Contributions
The author contributions are as follows: Mi Chen and Maurizio Taramasso composed the manuscript and produced the videos. Aris Moschovitis provided visions for significant intellectual content.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Disclosure
M.C. received grant from Boston Scientific and Edwards Lifesciences; has served as a consultant for Jenscare Scientific. M.T. has served as a consultant for Abbott Vascular, Boston Scientific, Edwards Lifesciences, Medtronic, PiCardia, CoreQuest, Shenqi Medical, Ventrimend, HiD Medical, Simulands, MEDIRA and CoreMedic; A.M. is consultant for Abbott Vascular.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Chen, M., Moschovitis, A. & Taramasso, M. Pacemaker lead dislocation during TricValve procedure with an extremely small superior vena cava. Int J Cardiovasc Imaging (2024). https://doi.org/10.1007/s10554-024-03067-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10554-024-03067-1