Monolithic HV-CMOS sensors for a beam monitoring system of therapeutic ion beams

, , , , , , , and

Published 19 March 2024 © 2024 The Author(s)
, , Citation B. Topko et al 2024 JINST 19 C03043 DOI 10.1088/1748-0221/19/03/C03043

1748-0221/19/03/C03043

Abstract

Nowadays, cancer treatment with ion beam is well established and studied. This method allows to deposit the maximum dose to the tumor and minimize the damage to healthy tissue, due to the Bragg peak of the ion energy deposition near the end of the particle range. During the treatment, it is possible to provide volumetric dose delivery by changing the particle energy (penetration depth) and adjusting the beam position via a magnetic system. For the beam monitoring system, the precise measurement of the beam direction, shape and fluence in real time becomes crucial to provide effective and safe dose delivery to the tumor. Additionally, the system should work for beam intensities up to 1010 s-1 for protons, be tolerant to 1 MeV neutron equivalent fluences up to 1015 cm-2 per year and be to tolerant to magnetic fields (for MR-guided ion beam). The studies presented in this article are focused on the application of the HitPix sensor family with counting electronics and frame-based readout for such a beam monitoring system. The HitPix sensors are monolithic pixelated silicon sensors based on HV-CMOS technology and have been developed at the ASIC and Detector Lab (ADL, KIT). Recent measurements with ion beams and a multi-sensor readout as well as future developments are discussed.

Export citation and abstract BibTeX RIS

Published by IOP Publishing Ltd on behalf of Sissa Medialab. Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.